<img height="1" width="1" style="display:none" src="https://www.facebook.com/tr?id=875423625897521&amp;ev=PageView&amp;noscript=1">

Sensing Savvy

Fistulography, Intra-access Blood Flow Measurement Correlations

Posted by Deborah Brouwer-Maier RN, CNN on Jun 12, 2017 6:30:00 AM

intra access blood flow measurementsThe success of percutaneous transluminal angioplasty (PTA) in stenotic hemodialysis prosthetic grafts and autogenous fistulas is traditionally assessed by the post-intervention vessel diameter of the grafts or fistulas. Interventionalists at the University of San Diego sought to compare this traditional measure of success (vessel diameter), and pre- and post-intervention real-time intra-access blood flow rates with access blood flow measured during hemodialysis.

Radiologic images and intra-access blood flow measurements of 35 patients with 76 procedures were analyzed. Nineteen of the patients had autologous fistulas and 16 had prosthetic grafts. Access blood flow rates were also included, if they took place within five weeks after the intervention.

The study found that there were significant correlations between:

  • Initial access blood flow measured during hemodialysis and pre-intervention intra-access flow measurements as well as fistulography;
  • Post-intervention intra-access flow measurements and access blood flow during hemodialysis;

Final intra-access flow measurements and access blood flow correlated better than final fistulography and access blood flow. The clinicians also found no significant correlation between access survival and intra-access measurements after the intervention.

From their study, the interventionalists concluded that post-intervention intra-access blood flow measurements correlate more strongly than fistulography with access blood flow rates performed during hemodialysis with gold standard flow/dilution technology. They also cautioned that more studies are needed to determine the optimum target flow value to be achieved by a percutaneous intervention.

Download the Hemodialysis Handbook

Topics: Blood Flow Measurement